Operating Instructions
Microhmmeter Type DO5000
26
OTHER ISSUES
Due to inductive effects, ‘Overange Error’, ‘Polarity Error’, ‘Excess Resistance in Current Leads
Error’ and ‘Open Circuit Error’ will all be ignored until the final measurement.
If the Autorange changes the range during a measurement, then the ‘Number of Readings’
counter will reset to one.
If the measurement does not settle within the maximum number of readings, then the
‘Resistance’ bit (Bit 9) of the Questionable Data Condition/Event Registers will be set.
When in Manual mode, a short beep will be heard when the measurement is completed, or a
longer beep if the measurement failed to settle.